Skip to content
Recall Observatory FDA recall evidence

Device product

Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a variety of gas concentrations and alert clinical personnel when the concentration of anesthetic agent(s), oxygen, carbon dioxide or nitrous oxide moves outside of user-defined limits.

Z-2077-2014

June 18, 2014

Class II

Product summary

Firm
Spacelabs Healthcare Inc
Event
Event 68611
Status
Terminated
Classification
Class II
Quantity
8 units total (3 units in US)
Official record key
device-enforcement:Z-2077-2014

Official wording

Reason: With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.

Code information: Serial Numbers: 2518-000183 2518-000184 2518-000185 2518-000186 2518-000206 2518-000207 2518-000208 2518-000209

Distribution pattern: Worldwide Distribution - US in the state of Georgia and in the countries of : Malaysia, Paraguay, and France.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.