Skip to content
Recall Observatory FDA recall evidence

Device product

LifeShield LATEX-FREE HemoSet 100 mL Burette I.V. PlumSet, Nonvented, 105 inch with Prepierced Reseals, OPTION-LOK and Inline Dual Channel Cassette. For administration of blood and blood bags for use with Plum Series Infusers.

Z-1479-2014

February 25, 2014

Class II

Product summary

Firm
Hospira Inc.
Event
Event 67722
Status
Terminated
Classification
Class II
Quantity
3,893 sets
Official record key
device-enforcement:Z-1479-2014

Official wording

Reason: Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.

Code information: Product List Number: 112410403; Lot Numbers: 280055H, 341005H

Distribution pattern: US Distribution including the states of FL, IA,IN, KS, KY, ME, MO, MT, PA, TN, TX, UT, VA and WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Hospira identified an incorrect lower lid (set component) had been supplied and was being used during the manufacturing process of list number 11241-03, a HemoSet containing a burette and blood filter, intended for infusion of blood/blood products. In a gravity delivery, the correct lower lid dispenses 15 drops per mL and the incorrect lower lid found dispenses 10 drops per mL.

Field note

Send feedback

We'll only use this to respond to your feedback.