Skip to content
Recall Observatory FDA recall evidence

Device product

List No. 14217-28; Y-TYPE BLOOD SET; 200 Micron Filter, Prepierced Y-Site, Non-Vented, 80 Inch, Non-DEHP; For Blood Product Administration; Made in Costa Rica; Product Inquiries should be directed to Hospira., Inc., Lake Forrest, IL 60045 USA Hospira Blood sets are designed for Blood Product Administration

Z-1857-2013

April 01, 2013

Class I

Product summary

Firm
Hospira Inc.
Event
Event 65689
Status
Terminated
Classification
Class I
Quantity
1,104 units
Official record key
device-enforcement:Z-1857-2013

Official wording

Reason: It was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.

Code information: List Number: 14217-28; Lot Number: 962255H

Distribution pattern: Worldwide Distribution - USA (nationwide) and Puerto Rico and Internationally to Costa Rica.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It was discovered that there was a potential for the piercing pin on certain Hospira blood sets to puncture the outer wall of blood bags during insertion of the pin into the blood bag.

Field note

Send feedback

We'll only use this to respond to your feedback.