Skip to content
Recall Observatory FDA recall evidence

Device product

Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Z-1440-2014

March 05, 2014

Class I

Product summary

Firm
Draeger Medical Systems, Inc.
Event
Event 67677
Status
Terminated
Classification
Class I
Quantity
3
Official record key
device-enforcement:Z-1440-2014

Official wording

Reason: The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

Code information: Oxylog 3000 plus, Part number: 5704833 with serial numbers: ASEK-0003, ASEK-0004, ASEK-0005

Distribution pattern: USA Nationwide Distribution in the states of WI and NY.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.