Skip to content
Recall Observatory FDA recall evidence

Device product

Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.

Z-0402-2013

November 17, 2011

Class II

Product summary

Firm
Kerr Corporation
Event
Event 63512
Status
Terminated
Classification
Class II
Quantity
1330 units
Official record key
device-enforcement:Z-0402-2013

Official wording

Reason: The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.

Code information: Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System has no expiration date. The affected units were produced and distributed since February 2011.

Distribution pattern: Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, and Europe.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.