Sensus Healthcare Inc, SRT-100 Vision IPX 0, HFUS Module IPX 1, with software version 1.9.2; Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities

Z-1235-2023

February 10, 2023

Class II

Product summary

Firm
Sensus Healthcare, Inc.
Event
Event 91795
Status
Terminated
Classification
Class II
Quantity
23 units
Official record key
device-enforcement:Z-1235-2023

Official wording

Reason: When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.

Code information: Software version 1.9.2 loaded onto Serial Numbers: 1512-2002; 1512-2003; 1606-2007; 1609-2010; 1609-2011; 1609-2018; 1612-2023; 1612-2025; 1705-2032; 1709-2046; 1803-2068; 1906-2137; 1909-2149; 1912-2167; 2003-2177; 2112-2201; 2112-2202; 2112-2208; 2112-2218; 2203-2232; 2203-2241; 2206-2264; 2206-2272

Distribution pattern: US Distribution to states of: RX, IN, NM, FL, NY, FA, NC, IA, and IL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When the user pauses the beam delivery, the dose timer is reset (to zero), and as a result may deliver an unexpected dose to the patient.