The STERIS 4085 General Surgical Table Product Usage: is an electro-hydraulically operated surgical table designed to support all general surgical procedures.

Z-2026-2014

June 27, 2014

Class II

Product summary

Firm
Steris Corporation
Event
Event 68678
Status
Terminated
Classification
Class II
Quantity
3689
Official record key
device-enforcement:Z-2026-2014

Official wording

Reason: STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.

Code information: All Model Numbers

Distribution pattern: Worldwide Distribution - US (nationwide in the states of AL, AK, AZ, AR, CA,CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY in the countries of Canada, India, Vietnam, China, India, Korea, Mexico, costa Rica, Brazil, Taiwan, Australia, New Zealand, El Salvador, Indonesia, Hong Kong, Singapore, Nepal.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    STERIS engineering analysis has determined the D1 pressure switch in the hydraulic column manifold does not adequately detect pressure level. If one or more of the table floor lock feet become unlocked due to the loss of hydraulic pressure, there is a potential for the table hand control to incorrectly indicate that the floor lock feet are locked.