Skip to content
Recall Observatory FDA recall evidence

Device product

Stryker Color Cuff Non-Sterile Disposable Tourniquet Cuff -NS DISP 34X4,1BLA,2PRT QUICK Catalog number: 5921-034-235NS Indicated for use in surgical procedures that require the temporary occlusion of blood flow in a patient s extremities

Z-1342-2023

March 01, 2023

Class II

Product summary

Firm
Stryker Corporation
Event
Event 91849
Status
Ongoing
Classification
Class II
Quantity
1617 units (packs of 10)
Official record key
device-enforcement:Z-1342-2023

Official wording

Reason: Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications

Code information: GTIN: 07613327606188 Lot Numbers: 2022111806 2022112802 2022120501 2022120502 2022120503 2022120504 2022120505 2022120701

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Not able to achieve desired pressure and not holding pressure result from two different causes which have been reported to occur during initial set-up of the cuff onto the patient, may lead to operative site blood loss, hemorrhage or, in exceedingly rare situations, other major complications