Skip to content
Recall Observatory FDA recall evidence

Device product

Justright Surgical, JustRight 5mm Reload, REFJR-REL25-2.0, Sterile, EO, Rx Only It is a replacement cartridge used when more than one firing is required during a surgery.

Z-2691-2014

August 27, 2014

Class II

Product summary

Firm
Justright Surgical, LLC
Event
Event 69140
Status
Terminated
Classification
Class II
Quantity
3 cases containing 36 units
Official record key
device-enforcement:Z-2691-2014

Official wording

Reason: JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.

Code information: Lot: 75FD0417

Distribution pattern: Distributed in the states of Colorado, California and New York.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.

Field note

Send feedback

We'll only use this to respond to your feedback.