Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.

Z-1701-2014

May 13, 2014

Class II

Product summary

Firm
Hologic, Inc.
Event
Event 68298
Status
Terminated
Classification
Class II
Quantity
3 units
Official record key
device-enforcement:Z-1701-2014

Official wording

Reason: Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

Code information: Serial Numbers: 1800314 7263, 180031472, 18002023051PR

Distribution pattern: Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.