CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. It is not intended for homecare use.

Z-2006-2012

June 19, 2012

Class I

Product summary

Firm
CareFusion 203, Inc.
Event
Event 62349
Status
Terminated
Classification
Class I
Quantity
266
Official record key
device-enforcement:Z-2006-2012

Official wording

Reason: CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault. The unit will activate either a Low PEEP or Patient Circuit Fault Alarm both audibly and Visually to alert the healthcare professional. Ventilation delivery to the patient may be compromised.

Code information: serial numbers: 26562, 43634, 43642, 43643, 44536, 44537, 44540, 44541, 44542, 44543, 44544, 44545, 44546, 44552, 44553, 44554, 44555, 44557, 44558, 44559, 44560, 44561, 44562, 44563, 44564, 44565, 45318, 45320, 45321, 45322, 45323, 45324, 45325, 45326, 45327, 45328, 45329, 45330, 45333, 45334, 45335, 45336, 45337, 45338, 45339, 45340, 45341, 45342, 45343, 45344, 45345, 45346, 45735, 45736, 45737, 45738, 45739, 45740, 45741, 45742, 45743, 45744, 45745, 45746, 45748, 45750, 45751, 45752, 45753, 45754, 45902, 45903, 45904, 45905, 45906, 45907, 45908, 45909, 45910, 45911, 46038, 46039, 46040, 46041, 46042, 46043, 46044, 46045, 46046, 46047, 46048, 46049, 46050, 46051, 46052, 46632, 46634, 46635, 46636, 46637, 46638, 46639, 46640, 46641, 46642, 46643, 46644, 46645, 47069, 47070, 47071, 47073, 47074, 47075, 47076, 47077, 47078, 47079, 47080, 47081, 47082, 47083, 47210, 47211, 47212, 47213, 47214, 47215, 47216, 47217, 47219, 47221, 47222, 47223, 47225, 47226, 47227, 47229, 47230, 47231, 47232, 47233, 47234, 47236, 47237, 47238, 47239, 47538, 47539, 47540, 47541, 47542, 47543, 47544, 47545, 47546, 47547, 47548, 47549, 47550, 47551, 47552, 47553, 47554, 47555, 47556, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47565, 47566, 47567, 48759, 48760, 48761, 48762, 48763, 48764, 48765, 48766, 48767, 48768, 48769, 48770, 48771, 48772, 48773, 48774, 48775, 48776, 48777, 48778, 48779, 48780, 48781, 48782, 48783, 48784, 48785, 48786, 48787, 48788, 48868, 48869, 48870, 48871, 48872, 48873, 48874, 48875, 48876, 48878, 48881, 48883, 48884, 48885, 48886, 48887, 48889, 48890, 48891, 48892, 48893, 48894, 48895, 48897, 50390, 50391, 50392, 50393, 50394, 50395, 50397, 50409, 50411, 50412, 50416, 50417, 50418, 50419, 50648, 50650, 50651, 50652, 50654, 50656, 50657, 50658, 50659, 50660, 50661, 50662, 50663, 50664, 50665, 50666, 50667, 50668, 50672, 51569, 51574

Distribution pattern: Nationwide Distribution including AZ, CA, CO, FL, IL, KY, ME, MI, MO, MN, MT, ND, NJ, NM, NV, NY, OK, PA, SC, SD, TX , UT, WI, and WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault. The unit will activate either a Low PEEP or Patient Circuit Fault Alarm both audibly and Visually to alert the healthcare professional. Ventilation delivery to the patient may be compromised.