Skip to content
Recall Observatory FDA recall evidence

Device product

Apollo XT Suture Anchor, 5.5 x 15 Product Usage: Apollo XT Suture Anchors are intended for: Shoulder - Rotator Cuff Repair - Bankart Repair - SLAP Lesion Repair - Biceps Tenodesis - Acromio-Clavicular Separation Repair - Deltoid Repair - Capisular Shift or Capsulolabral Reconstruction Foot/Ankle - Lateral Stabilization - Medial Stabilization - Achilles Tendon Repair Knee - Medial Collateral Ligament Repair - Lateral Collateral Ligament Repair - Posterior Oblique Ligament Repair - Illiotibial Band Tenodesis Elbow - Biceps Tendon Reattachment - Ulnar or Radial Collateral Ligament Reconstruction Hip - Capsular Repair - Acetabular Labral Repair

Z-2430-2018

January 22, 2018

Class II

Product summary

Firm
Valeris Medical, LLC
Event
Event 80069
Status
Terminated
Classification
Class II
Quantity
113 units
Official record key
device-enforcement:Z-2430-2018

Official wording

Reason: The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.

Code information: Model No. XT-5515, Lot No. 2358, 2362, 2363

Distribution pattern: US Nationwide Distribution in the states of California and Oklahoma.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.

Field note

Send feedback

We'll only use this to respond to your feedback.