Foam electrodes with Silver-Carbon construction: P/N 202-9271, Cat No. 400-898, 2" x 2" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9281, Cat No. 202-9281, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9251, Cat No. 400-893, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9291, Cat No. 400-892, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9371, Cat No. 400-888, 2" x 3.5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9491, Cat No. 400-869, 1.25" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9511, Cat No. 400-853, 3" x 5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1541, Cat No. 300-154, 2" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1561, Cat No. 300-156, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1611, Cat No. 300-161, 4" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010, 22818 P/N 300-1631, Cat No. 300-163, 3" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8701, Cat No. 400-870, 2" x 4" Oval Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Z-0818-2013

November 30, 2009

Class II

Product summary

Firm
NAImco Inc dba Richmar Inc
Event
Event 63927
Status
Terminated
Classification
Class II
Quantity
56980 electrodes
Official record key
device-enforcement:Z-0818-2013

Official wording

Reason: Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

Code information: All Serial Numbers distributed between 11/01/2009 and 01/31/2010

Distribution pattern: Distributed in the states of GA and TN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.