Skip to content
Recall Observatory FDA recall evidence

Device product

BIOGRAPH mCT S(20) - 4R, MATERIAL NUMBER 10528958 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

Z-2000-2016

April 25, 2016

Class II

Product summary

Firm
Siemens Medical Solutions USA Inc.
Event
Event 74239
Status
Terminated
Classification
Class II
Quantity
1 unit
Official record key
device-enforcement:Z-2000-2016

Official wording

Reason: The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.

Code information: Serial Number 11022

Distribution pattern: Worldwide Distribution - US Nationwide and the countries of China and Malaysia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.