Skip to content
Recall Observatory FDA recall evidence

Device product

Product 28 consists of all product under product code: JDI and same usage: Item no: 32902605000 FEMORAL HD 28MM MEDIUM 32902604000 FEMORAL HD 28MM SHORT 32902604035 FEMORAL HD 28MM SHT+ 3.5M For use in total or hemi hip arthroplasty

Z-1126-2016

January 11, 2016

Class II

Product summary

Firm
Zimmer Manufacturing B.V.
Event
Event 72900
Status
Terminated
Classification
Class II
Quantity
0
Official record key
device-enforcement:Z-1126-2016

Official wording

Reason: As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.

Code information: lot no.: 62185968 62273370 62339644 62162469 62279774 62279774N 62221051 62279775 62279775N

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.