Skip to content
Recall Observatory FDA recall evidence

Device product

Part #313.221, 1.5mm Stardrive Screwdriver Bld Slf-Retain Shrt/42mm MQC, lot # 4341033, 4589566, 4703774, 4363781, 4589567, 4793575, 4904305, 4427700,4589568, 4830581, 4532370, 4673943, 4871657, mfg. 2/12/2002-5/23/2005 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.

Z-2196-2016

May 31, 2016

Class II

Product summary

Firm
Synthes (USA) Products LLC
Event
Event 74416
Status
Terminated
Classification
Class II
Quantity
383 units
Official record key
device-enforcement:Z-2196-2016

Official wording

Reason: DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.

Code information: Part #313.221, 1.5mm Stardrive Screwdriver Bld Slf-Retain Shrt/42mm MQC, lot # 4341033, 4589566, 4703774, 4363781, 4589567, 4793575, 4904305, 4427700,4589568, 4830581, 4532370, 4673943, 4871657, mfg. 2/12/2002-5/23/2005

Distribution pattern: US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.

Field note

Send feedback

We'll only use this to respond to your feedback.