Skip to content
Recall Observatory FDA recall evidence

Device product

Part # SD313.011, Matrix Neuro Screwdriver Blade MQC/MR Conditional/Short, Lot # UR75220, mfg 7/18/2007 Product Usage: The DePuy Synthes MR Safe Instruments and MR conditional Instruments are used to assist with the insertion or removal of an implant.

Z-2191-2016

May 31, 2016

Class II

Product summary

Firm
Synthes (USA) Products LLC
Event
Event 74416
Status
Terminated
Classification
Class II
Quantity
18 units
Official record key
device-enforcement:Z-2191-2016

Official wording

Reason: DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.

Code information: Part # SD313.011, Matrix Neuro Screwdriver Blade MQC/MR Conditional/Short, Lot # UR75220, mfg 7/18/2007

Distribution pattern: US Nationwide Distribution in the states of: AR,CA,CO,CT,DC,ID,IL,KY,LA,MA,MD,MN,MO,MS,MT,NY,PA and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    DePuy Synthes is voluntarily initiating this recall due to affected devices that were distributed, etched and or labeled with MR Safety information not meeting the current ASTM standard.

Field note