Skip to content
Recall Observatory FDA recall evidence

Device product

Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.

Z-2116-2017

February 15, 2012

Class III

Product summary

Firm
Medtest Holdings, Inc.
Event
Event 76985
Status
Terminated
Classification
Class III
Quantity
119.168L across All lots
Official record key
device-enforcement:Z-2116-2017

Official wording

Reason: A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.

Code information: Catalog numbers: HA461, HA761 , HA961 and lot codes:022204, 034801, 102401, 119503, and 134101.

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.

Field note

Send feedback

We'll only use this to respond to your feedback.