Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

Z-1651-2021

March 25, 2021

Class I

Product summary

Firm
The Biomed Guys
Event
Event 87768
Status
Ongoing
Classification
Class I
Quantity
22 units
Official record key
device-enforcement:Z-1651-2021

Official wording

Reason: Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.

Code information: Serial numbers of serviced devices, serviced 09/2020-02-2021: 12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,

Distribution pattern: Us distribution to the states of IN and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.