Device product

Draeger Infinity Acute Care System Monitoring Solution

Z-0714-2014

December 10, 2013

Class II

Product summary

Firm
Draeger Medical Systems, Inc.
Event
Event 67056
Status
Terminated
Classification
Class II
Quantity
439
Official record key
device-enforcement:Z-0714-2014

Official wording

Reason: After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.

Code information: with software versions VG2.0.3 and higher.

Distribution pattern: US Distribution including the states of: MA, NH, IN, CA, TX, NY, OK, and IL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased to reach the red alarm limit.