Skip to content
Recall Observatory FDA recall evidence

Device product

Stryker AVS Navigator Trials 6mm-15mm Manufactured by: Stryker Spine SAS Z.1 Marticot-33610-Cestas -France +33 (0) 5.57.97.06.30 http;//www.stryker.com Distributed in the USA by Stryker Spine 2 Pearl Court. Allendale NJ 07401-1677 USA +1-201-780-8000 Cestas France 33610. Intended Use: The trials are designed and intended to measure the height of the disc space intra-operatively.

Z-2033-2012

January 13, 2012

Class II

Product summary

Firm
Stryker Spine
Event
Event 61723
Status
Terminated
Classification
Class II
Quantity
273 units
Official record key
device-enforcement:Z-2033-2012

Official wording

Reason: Beginning in February 2011, Stryker Spine began receiving reports related to the shaft of the AVS Navigator trial failing at the junction of the trial head.

Code information: Class I; 510(k) Exempt Description: AVS¿ Navigator Trials. Size 6mm -15mm. Catalog Number: 48392206,48392207,48392208,48392209,48392210, 48392211,48392212,48392213,48392214,48392215. Lot Code: 104864,104866, 104867, 104868, 104869, 104870, 10E331, 104871, 104872, 104873, 104874, 104875.

Distribution pattern: Nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Beginning in February 2011, Stryker Spine began receiving reports related to the shaft of the AVS Navigator trial failing at the junction of the trial head.