Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens Mobilett XP Digital Mobile X-ray System Product Usage: The Mobilett XP and Mobilett XP Hybrid are radiographic systems designed for use in wards, intensive care, and premature birth-wards, pediatric and emergency departments, operating theatres as well as the central X-ray department.

Z-1985-2014

May 09, 2014

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 68382
Status
Terminated
Classification
Class II
Quantity
10
Official record key
device-enforcement:Z-1985-2014

Official wording

Reason: There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.

Code information: model no. 1818447, with serial nos.: 3284 3449 1026 1025 1010 1041 1054 2002 2058 2022

Distribution pattern: US Nationwide Distributions in the states of MO, GA, FL, NJ, CA, PA and including NM.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential fire hazard of certain Lithium-Ion batteries in the computers used in the original manufacturing or as spare parts between October 2010 and April 2011 for the Siemens Mobilett XP Digital mobile X-ray System, which might occur at any time with higher probability while the system is being charged and connected to the main power supply.

Field note

Send feedback

We'll only use this to respond to your feedback.