Device product

Xario Diagnostic Ultrasound System SSA-660A Used for routine diagnostic imaging

Z-0119-2014

August 05, 2013

Class II

Product summary

Firm
Toshiba American Medical Systems Inc
Event
Event 66059
Status
Terminated
Classification
Class II
Quantity
17 (1352 Total)
Official record key
device-enforcement:Z-0119-2014

Official wording

Reason: The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode. When the system is used under both conditions listed below, the temperature of the transducer head connected to the system may increase in temperature and possibly

Code information: Lot Numbers not utilized. Serial Numbers: LGA0792628 LGA0792564 LGA0792565 LGA0792531 LGA0792540 LGA0792541 LGA0792617 LGA0792554 LGA0792626 LGA0792627 LGA0792552 LGA0792616 LGA0792563 LGA0792553 LGA0792615 LGA0792551 LGA0792613

Distribution pattern: Nationwide Distribution including and US Territory Puerto Rico

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode. When the system is used under both conditions listed below, the temperature of the transducer head connected to the system may increase in temperature and possibly