Skip to content
Recall Observatory FDA recall evidence

Device product

ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.

Z-1819-2015

May 11, 2015

Class II

Product summary

Firm
Orthofix, Inc
Event
Event 71262
Status
Terminated
Classification
Class II
Quantity
36 units
Official record key
device-enforcement:Z-1819-2015

Official wording

Reason: There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.

Code information: All lot codes

Distribution pattern: Nationwide Distribution including CA, VA, FL, ID, CO, OR, PA, OK, SC, and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.