Device product

AIA-360 Analyzer is capable of performing two methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay. An antigen-antibody reaction begins by combining a patient sample, control, or calibrator with a diluent in an immunoreaction test cup from the AIA-PACK reagent series.

Z-0613-2019

March 22, 2017

Class II

Product summary

Firm
Tosoh Bioscience Inc
Event
Event 81170
Status
Terminated
Classification
Class II
Quantity
438
Official record key
device-enforcement:Z-0613-2019

Official wording

Reason: The firm has become aware of a potential issue with the AIA-360 analyzer that could prevent the table home sensor from detecting the table home position. If the malfunction with the table home sensor occurs on the AIA-360, the sensor board may need to be replaced. If the home sensor fails to detect the home position, the AIA-360 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received three (3) complaints related to this issue with no serious injuries reported.

Code information: All units with serial numbers prior to "1701"

Distribution pattern: Nationwide distribution. International distribution to Honduras, Caribbean, Chile, Colombia, Costa Rica, Uruguay, Ecuador, Panama, Mexico, Canada, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm has become aware of a potential issue with the AIA-360 analyzer that could prevent the table home sensor from detecting the table home position. If the malfunction with the table home sensor occurs on the AIA-360, the sensor board may need to be replaced. If the home sensor fails to detect the home position, the AIA-360 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received three (3) complaints related to this issue with no serious injuries reported.