Skip to content
Recall Observatory FDA recall evidence

Device product

Justright Surgical, JustRight 5mm Stapler, REF JR-ST25-2.0, Sterile, EO, Rx Only, JR-ST25-2.0 product is packaged in cases of 6. The JustRight 5mm Stapler is intended for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection, and creation of anastomosis.

Z-2690-2014

August 27, 2014

Class II

Product summary

Firm
Justright Surgical, LLC
Event
Event 69140
Status
Terminated
Classification
Class II
Quantity
3 cases containing 18 units
Official record key
device-enforcement:Z-2690-2014

Official wording

Reason: JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.

Code information: Lot: 75DD0406

Distribution pattern: Distributed in the states of Colorado, California and New York.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    JustRight Surgical is recalling specific lots of JustRight 5mm Stapler and Reload products because they may deliver poor distal staple formation, past the end of the knife cut.