Device product
PERSONA The Personalized Knee System CR Tibial Articular Surface Provisional with the following: Left Size 3-9 CD Top; Left Size 3-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Left Size 3-9 CD Top; Left Size 3-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.
Z-2577-2014
Product summary
- Firm
- Zimmer, Inc.
- Event
- Event 68878
- Status
- Terminated
- Classification
- Class II
- Quantity
- 77,737 total units
- Official record key
device-enforcement:Z-2577-2014
Official wording
Reason: Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.
Code information: Part Numbers: 42517000410, 42517000510, 42517000610, 42517000710, 42527000410, 42527000510, 42527000610, and 42527000710. Lot numbers: 62024975, 62043546, 62144792, 62152725, 62165538, 62187318, 62337961, 62372407, 62386776, 62397159, 62405917, 62456263, 62499170, 62551386, 62565051, 62568272, 62572572, 62663168, 62691384, 62707494, 62717346, 62020920, 62144793, 62156313, 62165539, 62184228, 62353932, 62356176, 62385100, 62420242, 62468277, 62538412, 62564340, 62611088, 62612263, 62690056, 62702295, 62717337, 62736078, 62024976, 62054776, 62147232, 62177243, 62187319, 62187320, 62359057, 62364955, 62391222, 62414006, 62461233, 62558583, 62564345, 62608889, 62638657, 62646286, 62702749, 62724348, 62394279, 62443397, 62487490, 62492802, 62506001, 62563915, 62568271, 62704633, 62055686, 62144796, 62161537, 62162597, 62256286, 62351297, 62356578, 62385099, 62404282, 62424802, 62455897, 62538392, 62565037, 62595855, 62611083, 62663162, 62680121, 62686205, 62695286, 62724347, 62724744, 62019863, 62156317, 62165548, 62168386, 62184268, 62256288, 62344258, 62356586, 62391232, 62391233, 62455892, 62504877, 62551337, 62564335, 62597816, 62638685, 62646284, 62695878, 62036673, 62138318, 62149715, 62189326, 62189327, 62356593, 62377738, 62391236, 62420225, 62461232, 62564360, 62564361, 62595865, 62611086, 62680120, 62690951, 62710615, 62761109, 62394280, 62428338, 62444959, 62500931, and 62568278.
Distribution pattern: Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of: AUSTRALIA, INDIA, JAPAN, KOREA, SINGAPORE, AUSTRIA, BELGIUM, FRANCE, GERMANY, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWITZERLAND, UNITED KINGDOM, and UNITED ARAB EMIRATES.
Derived failure modes
-
Unknown
Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.