Skip to content
Recall Observatory FDA recall evidence

Device product

Irix-C Forceps Inserter; Product Code/UDI: T066-0059/ M697T06600591

Z-2500-2020

May 12, 2020

Class II

Product summary

Firm
XTANT Medical
Event
Event 85789
Status
Terminated
Classification
Class II
Quantity
4 units
Official record key
device-enforcement:Z-2500-2020

Official wording

Reason: The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.

Code information: LOT: XS1920458A

Distribution pattern: AZ, TX and WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The strike plate fractured off from the forceps inserter devices with minimal impaction or force due to unsatisfactory weld.

Field note

Send feedback

We'll only use this to respond to your feedback.