Skip to content
Recall Observatory FDA recall evidence

Device product

t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.

Z-2471-2018

April 23, 2018

Class II

Product summary

Firm
Tandem Diabetes Care Inc
Event
Event 79887
Status
Ongoing
Classification
Class II
Quantity
55
Official record key
device-enforcement:Z-2471-2018

Official wording

Reason: The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.

Code information: Serial Numbers: 483731, 483930, 483992, 484013, 492055, 492368, 501157, 501177, 501178, 501187, 501191, 501200, 501202, 501205, 501207, 501268, 501285, 501364, 501368, 501370, 501376, 501385, 501386, 501387, 501432, 501435, 501440, 501443, 501444, 501445, 501466, 501479, 501488, 501490, 501491, 501495, 501503, 501508, 501509, 501519, 501522, 501523, 501526, 501535, 501560, 504522, 504531, 504535, 504540, 504753, 504972, 504990, 505025, 90472671, 90488584

Distribution pattern: U.S. Consignees: IN, AZ, UT, TX, NY, NJ, MD, MN, IL, CO, AL, WI, CA, PA, OH, OR, MS, FL, NM, ID, NV, KY, VA,

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways: A) by triggering a Malfunction 4 Alarm, or B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.

Field note

Send feedback

We'll only use this to respond to your feedback.