Skip to content
Recall Observatory FDA recall evidence

Device product

Vivo 45LS Ventilator- intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation Model Number: 230000 ¿ If the recalled product is a software, include version number: firmware versions 5.0.5 or earlier

Z-1382-2022

June 02, 2022

Class II

Product summary

Firm
Breas Medical, Inc.
Event
Event 90433
Status
Ongoing
Classification
Class II
Quantity
1,113 units
Official record key
device-enforcement:Z-1382-2022

Official wording

Reason: A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained

Code information: Unique Device Identifier (UDI): UDI 07321822300004. Serial number ranges in scope are: D******, K******, M01**** to M040124, where asterisk (*) denotes a digit from 0 to 9.

Distribution pattern: Nationwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A forced shutdown of one of the processors did not generate a watch dog alarm, would develop into a failure AND the instructions to monitor a ventilator dependent patient are NOT followed, the health consequences could potentially be Permanent impairment or life threatening if medical intervention is not obtained

Field note

Send feedback

We'll only use this to respond to your feedback.