Skip to content
Recall Observatory FDA recall evidence

Device product

MODEL: X-RAY TV SYSSTEM SONIALVISION safire17

Z-0869-2022

January 10, 2022

Class II

Product summary

Firm
Shimadzu Medical Systems
Event
Event 89596
Status
Ongoing
Classification
Class II
Quantity
12 systems
Official record key
device-enforcement:Z-0869-2022

Official wording

Reason: It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Code information: Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013

Distribution pattern: U.S.: AK, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, WA, and WV. O.U.S.: N/A

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.

Field note

Send feedback

We'll only use this to respond to your feedback.