Skip to content
Recall Observatory FDA recall evidence

Device product

BD BBL Sensi Disc Minocycline - 30 ug, used for semi-quantitative in vitro susceptibility testing. Catalog Number 231250 and 231251¿

Z-1034-2024

January 08, 2024

Class II

Product summary

Firm
Becton Dickinson & Co.
Event
Event 93789
Status
Ongoing
Classification
Class II
Quantity
2,363,168 total units
Official record key
device-enforcement:Z-1034-2024

Official wording

Reason: There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Code information: Catalog No. 231250 UDI-DI 00382902312505 Lots 1085656 1277053 1301694 1334451 2025108 2119580 2146396 3181025; Catalog No. 231251¿ UDI-DI 30382902312513 Lots 1007140 2146396 3181025 3234149 1085656 1210753 1277053 1301694 1334451 2025108 2119580

Distribution pattern: Domestic distribution nationwide. International distribution worldwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a delay in diagnosis, the selection of inappropriate antibiotics, or extended duration of antibiotic exposure and the treatment process.

Field note

Send feedback

We'll only use this to respond to your feedback.