Skip to content
Recall Observatory FDA recall evidence

Device product

DUR PCA MTK REV INS RT Intended for knee replacement Product Number: 6728-2-711

Z-0922-2024

December 01, 2023

Class II

Product summary

Firm
Howmedica Osteonics Corp.
Event
Event 93687
Status
Ongoing
Classification
Class II
Quantity
Unavailable
Official record key
device-enforcement:Z-0922-2024

Official wording

Reason: Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Code information: GTIN: 04546540322951 Lot Numbers: 52177701

Distribution pattern: Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Field note

Send feedback