Skip to content
Recall Observatory FDA recall evidence

Device product

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Z-0974-2024

January 06, 2022

Class II

Product summary

Firm
Philips North America
Event
Event 93737
Status
Ongoing
Classification
Class II
Quantity
2
Official record key
device-enforcement:Z-0974-2024

Official wording

Reason: Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Code information: UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030

Distribution pattern: Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

Field note

Send feedback

We'll only use this to respond to your feedback.