Skip to content
Recall Observatory FDA recall evidence

Device product

SCORPIO RECESSED PATELLA- Intended for Knee Replacement Part Number: 3044-0026

Z-0869-2024

December 01, 2023

Class II

Product summary

Firm
Howmedica Osteonics Corp.
Event
Event 93687
Status
Ongoing
Classification
Class II
Quantity
94 units US; 131 units OUS
Official record key
device-enforcement:Z-0869-2024

Official wording

Reason: Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Code information: GTIN: 07613327033779 Lot Numbers: PVDY TEN1 AMN1 V644 MHHE 4T1K V355 55LK JR6R

Distribution pattern: Nationwide Foreign: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, India, Japan, Korea, Latin America, Mexico, Poland, Sweden, Switzerland, Taiwan, UK, & Venlo (Netherlands). ***Updated 1/25/24*** Foreign distribution added: Austria, Finland, Germany, Ireland, Hong Kong, New Zealand, Norway, and Spain.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

Field note

Send feedback

We'll only use this to respond to your feedback.