Skip to content
Recall Observatory FDA recall evidence

Device product

Aeos Robotic Digital Microscope, Product Code: PV010

Z-2139-2024

May 13, 2024

Class II

Product summary

Firm
Aesculap Inc
Event
Event 94623
Status
Ongoing
Classification
Class II
Quantity
19 units
Official record key
device-enforcement:Z-2139-2024

Official wording

Reason: Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Code information: Product Code: PV010, UDI/DI: 04046955206742; Serial Numbers: 1011, 1005, 1057, 1032, 1038, 1054, 1036, 1060, 1056, 1089, 1093, 1027, 1061, 1004, 1026, 1035, 1097, 1084, 1063

Distribution pattern: US Nationwide distribution in the states of AR, FL, IL, IN, KY, MI, NC, NE, OK, SD, TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.

Field note

Send feedback

We'll only use this to respond to your feedback.