SlideView VS200 slide viewer,

Z-2416-2024

July 22, 2024

Class II

Product summary

Firm
EVIDENT SCIENTIFIC INC
Event
Event 95028
Status
Ongoing
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-2416-2024

Official wording

Reason: A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Code information: 65159 67105 67104 69153 67106 69154

Distribution pattern: US

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.