Skip to content
Recall Observatory FDA recall evidence

Device product

Posi-Stop Injection Needle, 25 Gauge, Catalog Number: 4704

Z-0473-2025

October 04, 2024

Class II

Product summary

Firm
Hobbs Medical, Inc.
Event
Event 95577
Status
Ongoing
Classification
Class II
Quantity
6 units
Official record key
device-enforcement:Z-0473-2025

Official wording

Reason: Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Code information: Catalog Number: 4704; UDI-DI: M84947040; Lot Numbers: H10-23-106, H11-23-059.

Distribution pattern: Worldwide - US Nationwide distribution in the states of CT, GA, IN, MI, MN, MS, NH, NJ, NY, OH, TN, VT, WA, WI, WV and the country of Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Injection solutions would not flow as intended through the tip of 25ga-sized needles. Investigations that some units contained debris from the manufacturing process within the inner sheath that blocked the ID of the needle.

Field note

Send feedback

We'll only use this to respond to your feedback.