Skip to content
Recall Observatory FDA recall evidence

Device product

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Z-0664-2025

May 31, 2023

Class II

Product summary

Firm
Micro-X Ltd.
Event
Event 95603
Status
Ongoing
Classification
Class II
Quantity
5
Official record key
device-enforcement:Z-0664-2025

Official wording

Reason: Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Code information: UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

Distribution pattern: Worldwide and US Nationwide Distribution: NJ, PR, CA

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Field note

Send feedback

We'll only use this to respond to your feedback.