Skip to content
Recall Observatory FDA recall evidence

Device product

VACUETTE SAFELINK, REF: 450210

Z-0802-2025

November 19, 2024

Class II

Product summary

Firm
Greiner Bio-One GmbH
Event
Event 95748
Status
Ongoing
Classification
Class II
Quantity
188,800 pcs
Official record key
device-enforcement:Z-0802-2025

Official wording

Reason: The VACUETTE SAFELINK Holder with male luer lock may leak blood due to the protective rubber sleeve bunching up and not retracting when using the device.

Code information: UDI-DI: 19120049207487, Lot: A240338Q, Expiration Date: 270215

Distribution pattern: US Nationwide distribution in the states of IL, PA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The VACUETTE SAFELINK Holder with male luer lock may leak blood due to the protective rubber sleeve bunching up and not retracting when using the device.