Skip to content
Recall Observatory FDA recall evidence

Device product

The FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.

Z-0805-2025

December 03, 2024

Class III

Product summary

Firm
Qualigen Inc
Event
Event 95807
Status
Ongoing
Classification
Class III
Quantity
426
Official record key
device-enforcement:Z-0805-2025

Official wording

Reason: Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.

Code information: UDI-DI: 20816467020345, Lot numbers: 2408018-1 Exp: 2025-09-05; 2409015-1, Exp: 2025-06-27; and 2409015-2, Exp: 2025-06-26.

Distribution pattern: Distribution was made to AR, CA, CO, FL, IL, IN, MA, MO, NC, NY, OH, OK, TN, TX, and VA. There was no military/government/foreign distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.

Field note

Send feedback

We'll only use this to respond to your feedback.