Skip to content
Recall Observatory FDA recall evidence

Device product

Vios Monitoring System Bedside Monitor Model BSM2050

Z-1292-2025

August 09, 2024

Class II

Product summary

Firm
Murata Vios, Inc.
Event
Event 96252
Status
Ongoing
Classification
Class II
Quantity
50
Official record key
device-enforcement:Z-1292-2025

Official wording

Reason: Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.

Code information: 1.56.01 UDI: +B595BSM20500/$$7V1.56.01/16D20230630S 1.57.20 UDI: +B595BSM20500/$$7V1.57.20/16D20231129Y Software Versions 1.56.01 and 1.57.20

Distribution pattern: US Nationwide distribution in the states of Florida, Louisiana, Maryland, Montana, New York, Ohio, and Texas.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Batteries may deplete over time while not in use if devices are stored without being plugged in to wall outlet. Depleted batteries may lead to tablet connectivity issues.

Field note

Send feedback

We'll only use this to respond to your feedback.