Skip to content
Recall Observatory FDA recall evidence

Device product

NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

Z-0029-2026

September 09, 2025

Class I

Product summary

Firm
NOXBOX LTD
Event
Event 97516
Status
Ongoing
Classification
Class I
Quantity
1667 units
Official record key
device-enforcement:Z-0029-2026

Official wording

Reason: The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

Code information: Model Numbers: (1) NOXBOXI (NBL (Manufacturer)), (2) REQNOXBOXI (Linde (U.S. Distributor)); UDI-DI: 05060541640009; All serial numbers.

Distribution pattern: Domestic: TN;

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

Field note

Send feedback

We'll only use this to respond to your feedback.