Skip to content
Recall Observatory FDA recall evidence

Device product

Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG

Z-1100-2026

December 19, 2025

Class II

Product summary

Firm
ICU Medical Inc.
Event
Event 98168
Status
Ongoing
Classification
Class II
Quantity
64290 total
Official record key
device-enforcement:Z-1100-2026

Official wording

Reason: Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.

Code information: 1. 4044EG-1 UDI-DI 30351688073062 Lots 6141992 6156228 2. 4080P-2 UDI-DI 30351688006176 Lots 6141999 6142000 6147906 3. 4084P-1 UDI-DI 30351688029540 Lots 6147907 (corrected 1/22/26, removed "-LSM" suffix) 4. 4080PE UDI-DI 30351688039150 Lot 6131839 5. 4080PG UDI-DI 30351688081173 Lots 6131838 6142001 6142002

Distribution pattern: US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.