Skip to content
Recall Observatory FDA recall evidence

Device product

BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers.

Z-0997-2026

October 14, 2025

Class II

Product summary

Firm
Beckman Coulter Ireland, Inc.
Event
Event 97842
Status
Ongoing
Classification
Class II
Quantity
36,981 reagents (U.S.-35,320, O.U.S.-1,661)
Official record key
device-enforcement:Z-0997-2026

Official wording

Reason: Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Code information: Catalog Numbers / UDI-DI codes: OSR6137 - 15099590020187 Lot Numbers: 2681 2682 2683 2684 2685 2686 2687 2688 2689 2690 2693 2694 2695 2696 OSR6237 - 15099590010348 Lot Numbers: 2702 2703 2704 2705 2706 2707 2708 2709 2710 2712 2713 2714 2715 2716 OSR6637 - 15099590060206 Lot Numbers: 2726 2728 2729 2730 2731 2732 2733 2734 2736 2737 2738 2739

Distribution pattern: Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guan, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Brazil, Canada, Ecuador, Ireland, Mexico, Netherlands, Panama, Taiwan, and Province of China.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Field note