Skip to content
Recall Observatory FDA recall evidence

Device product

Diversatek Healthcare Viper¿ 3-Stage Wire Guided Balloon Dilator, 12mm-13.5mm-15mm, REF 1205-15

Z-1939-2026

March 18, 2026

Class II

Product summary

Firm
Diversatek Healthcare
Event
Event 98649
Status
Ongoing
Classification
Class II
Quantity
527 units
Official record key
device-enforcement:Z-1939-2026

Official wording

Reason: The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

Code information: UDI/DI 00816734021781, Lot Numbers: 004758, 005124, 005309, 005629

Distribution pattern: US: MO, CA, FL, TX, AZ, PA, RI, NC, IL, MS, KS, NY, OH, VA, LA, IA, MI, NJ, AL, Brazil, Belgium

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The catheter inflation tag attached to the device may contain an incorrect part number with incorrect balloon diameter sizes and inflation pressures for each stage.

Field note

Send feedback

We'll only use this to respond to your feedback.