Skip to content
Recall Observatory FDA recall evidence

Device product

LIFEPAK 15, REF: 99577-000025, 99577-000047, 99577-001256, 99577-001217, 99577-001241, 99577-001255, 99577-001219, 99577-001957, 99577-001935, 99577-001964, 99577-001955, 99577-001956, 99577-001588, 99577-001958, 99577-001950, 99577-001962, 99577-002178, 99577-001372, 99577-001938, 99577-000045; LIFEPAK 1000, REF: 99425-000025, 99425-000023, 99425-000096; LIFEPAK 35, REF: 99335-000013, 99335-000031; LIFEPAK 20e, REF: 99507-000122

Z-2065-2026

April 13, 2026

Class II

Product summary

Firm
Physio-Control, Inc.
Event
Event 98418
Status
Ongoing
Classification
Class II
Quantity
206
Official record key
device-enforcement:Z-2065-2026

Official wording

Reason: Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

Code information: Device/UDI/Serial Number: LIFEPAK 15/0088387391164815011543166573, 0088387387169015033143394891, 0088387387169015050743524360, 0088387387169015092843914574, 0088387387204815093043915122, 0088387387169016010444187447, 0088387387162116012044252611, 0088387387169016012644262721, 0088387387169016012644277082, 0088387387169016012744282280, 0088387387171316060244574340, 0088387387171316080444896329, 0088387387169016083144962804, 0088387387170617013145447229, 0088387387169017041245743786, 0088387387171317050945834209, 0088387387171317062845998142, 0088387387171317082846103043, 0088387387169017101046303288, 0088387387171317103146303092, 0088387387162117122146587130, 0088387387171317122646602003, 0088387387171317122846608783, 0088387387171317122946614788, 0088387387171318022246813507, 0088387387169018032946966267, 0088387387171318061947337396, 0088387387169018070547401928, 0088387387171318092747611441, 0088387387169018112947784931, 0088387387169018112947786317, 0088387387171318121847847661, 0088387387171318121847848950, 0088387387708118121947845317, 0088387387708118121947855371, 0088387387169018122047857172, 0088387387169018122047858660, 0088387387171318122847872868, 0088387387171319031248018370, 0088387387169019061248247559, 0088387387168319062748298704, 0088387387169019100848453060, 0088387387171319101448455890, 0088387387171319121848586351, 0088387387171319121948586599, 0088387387168320012848651507, 0088387387171320030548701689, 0088387387708120042848784975, 0088387387171320071648887706, 0088387387168320081748921846, 0088387387169020092448981398, 0088387387169020102649032288, 0088387387169020102649050140, 0088387387162120110249055931, 0088387387168320122949156402, 0088387387169021011149162067, 0088387387169021011149167463, 0088387387169021013049212578, 0088387387169021020949225869, 0088387387169021020949226060, 0088387387176821021949240566, 0088387387168321033149345530, 0088387387169021042149378981, 0088387387169021042649381781, 0088387387169021042749381576, 0088387387168321081849559854, 0088387387168321081949560127, 0088387387164521091649607224, 0088387387171321102149662806, 0088387377697122070750045009, 0088387387169022090150136719, 0088387387171322090850152289, 0088387387708122091350158293, 0088387387169022101950223934, 0088387387708122112150288760, 0088387387171322120150307353, 0088387387169023062850629377, 0088387387169023072450653772, 0088387387169023072450667197, 0088387387708123072850669746, 0088387387708123082850714419, 0088387387171323082950720588, 0088387387169023112150850129, 0088387387708123122050883524, 0088387388434824052251092313, 0088387387170624121151422867, 0088387387170624122351440213, 0088387387170624122351444039, 0088387388434824052251092313, 0088387387169019100848453060, 0088387387201715090843839264, 0088387387171318121847850793/ 38523489, 39736196, 40465345, 40717931, 40765749, 41130420, 41139184, 41420681, 41446039, 41446693, 41704583, 42221131, 42920956, 42921954, 43166573, 43394891, 43524360, 43914574, 43915122, 44187447, 44252611, 44262721, 44277082, 44282280, 44574340, 44896329, 44962804, 45447229, 45743786, 45834209, 45998142, 46103043, 46303288, 46303092, 46587130, 46602003, 46608783, 46614788, 46813507, 46966267, 47337396, 47401928, 47611441, 47784931, 47786317, 47847661, 47848950, 47845317, 47855371, 47857172, 47858660, 47872868, 48018370, 48247559, 48298704, 48453060, 48455890, 48586351, 48586599, 48651507, 48701689, 48784975, 48887706, 48921846, 48981398, 49032288, 49050140, 49055931, 49156402, 49162067, 49167463, 49212578, 49225869, 49226060, 49240566, 49345530, 49378981, 49381781, 49381576, 49559854, 49560127, 49607224, 49662806, 50045009, 50136719, 50152289, 50158293, 50223934, 50288760, 50307353, 50629377, 50653772, 50667197, 50669746, 50714419, 50720588, 50850129, 50883524, 51092313, 51422867, 51440213, 51444039, 51092313, 48453060, 43839264, 47850793, 38369584. LIFEPAK 1000/0088507469887815020943235807, 0088507469895315031343346860, 0088507469887815111944068667, 0088507469887815111944068669, 0088507469895316011844239775, 0088507469887816041444539726, 0088507469887817061745966118, 0088507469887817061745966121, 0088507469887817061745966122, 0088507469887817061745966126, 0088507469887817061945966131, 0088507469887817062945994506, 0088507469887817062946004311, 0088507469887817063046012479, 0088507469887818060547291148, 0088387381546519080648366102, 0088387381545820092948987376, 0088387381545821011249174469, 0088387381545821081349544565, 0088387381545821090249586700, 0088387381546521122249752235, 0088387381545822012649794992, 0088387381545822032949882696, 0088387381545822032949882698, 0088387381545822032949882819/ 34933095, 35096556, 35184465, 36153943, 36199099, 36560529, 36560531, 37119097, 37226402, 37330295, 38074718, 38074722, 38074727, 38074737, 38074741, 38074744, 38074745, 38074747, 38074748, 38074749, 38077938, 38077939, 38077940, 38077941, 38077944, 38378802, 38513659, 38656117, 38656121, 38656122, 38957375, 38957392, 38957397, 38957399, 38957402, 38972629, 38972677, 39775405, 39775406, 39775407, 39775415, 39964908, 40323073, 40412729, 40541187, 40584085, 40594714, 40707896, 40725568, 40725569, 40725571, 40725572, 40725575, 40725584, 40725582, 40725583, 41306695, 41692754, 41863786, 42504017, 42520778, 42803174, 42921847, 43235807, 43346860, 44068667, 44068669, 44239775, 44539726, 45966118, 45966121, 45966122, 45966126, 45966131, 45994506, 46004311, 46012479, 47291148, 48366102, 48987376, 49174469, 49544565, 49586700, 49752235, 49794992, 49882696, 49882698, 49882819. LIFEPAK 35/0088387377265224042251041190, 0088387377265224041851042010, 0088387377265224042251044581, 0088387376560924071151156570, 0088387376560924071151157072, 0088387376560924071151157449, 0088387376560924080651182632, 0088387376560924091251263526, 0088387376560924121051417585, 0088387376560925020451496956/ 51041190, 51042010, 51044581, 51156570, 51157072, 51157449, 51182632, 51263526, 51417585, 51496956. LIFEPAK 20e/0088387385386320082448935344, 0088387385386320082448935408, 0088387385386321080249537075/ 48935344, 48935408, 49537075

Distribution pattern: US Nationwide distribution in the states of MI, WA, KY, IN, FL, NC, TX, LA, VA, NY, MA, CO, PA, OH, SC, OK, MN, WV, NJ, WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.

Field note

Send feedback

We'll only use this to respond to your feedback.