Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 99223

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 18, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Eleva with Flat Detector.

Z-2411-2026
Recall number
Z-2411-2026
Initiated
May 18, 2026
Classification
Class II
Status
Ongoing
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

Code information

All codes

Distribution pattern

US and ROW.

device · product 2 of 2

Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # 722010 for Allura Xper FD10; Model # 722011 for Allura Xper FD10/10; Model # 722012 for Allura Xper FD20; Model # 722013 for Allura Xper FD20 Biplane; Model # 722015 for Allura Xper FD20 OR Table; Model # 722019 for Allura Xper FD10/10 OR Table; Model # 722020 for Allura Xper FD20 Biplane OR Table; Model # 722022 for Allura Xper FD10 OR Table; Model # 722023 for Allura Xper FD20 OR Table; Model # 722025 for Allura Xper FD20 Biplane OR Table; Model # 722026 for Allura Xper FD10; Model # 722027 for Allura Xper FD10/10; Model # 722028 for Allura Xper FD20; Model # 722029 for Allura Xper FD20/10; Model #722033 for Allura Xper FD10 OR Table; Model #722035 for Allura Xper FD20 OR Table; Model #722038 for Allura Xper FD20/20; Model #722039 for Allura Xper FD20/20 OR Table; Model #722058 for Allura Xper FD20/15; Model #722059 for Allura Xper FD20/15 OR Table.

Z-2412-2026
Recall number
Z-2412-2026
Initiated
May 18, 2026
Classification
Class II
Status
Ongoing
Quantity
Unavailable

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

Code information

All codes

Distribution pattern

US and ROW.

Field note

Send feedback

We'll only use this to respond to your feedback.