Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
- Recall number
- Z-2405-2026
- Initiated
- May 12, 2026
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Abbott Medical
- Quantity
- 11182 programs
App-derived interpretation
Software malfunction
Official device-enrichment evidence · Sourced
Software Design Change
Inspect official wording and provenance
Reason for recall
Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
Code information
Programmer Model 3650; Software Model 3330; GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)
Distribution pattern
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet