Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 98892

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 07, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Fresenius Medical Care Holdings, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Bicarby Dialysate; Model number: RFP-400-G;

Z-2401-2026
Recall number
Z-2401-2026
Initiated
May 07, 2026
Classification
Class II
Status
Ongoing
Quantity
4752

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.

Code information

Model number: RFP-400-G; UDI-DI: 00840861102839; Lot codes: 25SG06026, 25SG06027;

Distribution pattern

Domestic: TX.

Field note

Send feedback

We'll only use this to respond to your feedback.